Microwave Circuits and
Metamaterials
Project 7
Overview
Remain in same project groups for the
semester.
The objective of this project is to measure third-order
nonlinearities and OIP3 in the lab.
NOTE: Use the Project Report Template and keep answers to questions on consecutive sheets
of paper with all plots at the end.
IN NO CASE may code or files be exchanged between students, and
each student must answer the questions themselves and do their own
plots, NO COPYING of any sort! Nevertheless, students are
encouraged to collaborate in the lab session.
Only turn in requested plots ( Pxx )
and requested answers to questions ( Qxx ).
Part 1
- In this part,
third-order nonlinearities are measured in the lab.
- Two different test stations are set up in the lab, each
station with different equipment and a different amplifier to be
measured
- Test Setup/Station 1
- The test setup for test station 1 is shown below
- The top instrument is an Agilent
N1996A spectrum analyzer
- The bottom instrument is an Agilent
E4438C Vector Signal Generator
- First, connect the cable directly from the signal generator
into the spectrum analyzer to check the frequencies and power
levels of the two input signal frequencies. (The Agilent E4438C Vector Signal
Generator is capable of internally creating 2 signals.)
- Second, set the signal generator for -25 dBm, such
that you observe two frequencies at 199.5 and 200.5 MHz, each
one at a power level of -24 dBm (24 dB below 1 milliwatt) on
the spectrum analyzer (red arrow below)
- If the signals are already present on the spectrum analyzer
screen, only minor adjustments should be needed.
If the instructor is not available, you may need to refer to
the detailed setup instructions at the end of this document
- Make sure that the signal generator is set for -25 dBm
- Then, you should observe a spectrum similar to the following
- Press "Marker" and "PeakSearch" to set a marker on the left
spectral line (red arrow above)
- The power level should be approximately -29 dBm at 199.5
MHz,
- Record this power level, Pin in dBm, and frequency (199.5
MHz) of the left peak (red arrow above). The power level that you record here will
be the input power level (Pin) to the amplifier that you will
measure later ( Q1 )
- Make sure that the marker readout is visible, and you can
clearly see the power level and frequency of the peak at the
red arrow above, then take a photograph for your report ( P1 )
- Connect the MAR-1 amplifier as shown below, with the
spectrum analyzer connected to the output and with the signal
generator connected to the input and 10Vdc power
(red=positive, black=negative)
- Recheck to make sure that the signal generator is set for
-25 dBm
- You should observe an output spectrum similar to the
following
- Make sure that the marker readout is visible, and you can
clearly see the power level and frequency of the peak at the
red arrow above, then take a photograph for your report ( P2 )
- Record the power level, Pout in dBm, and frequency (199.5
MHz) of the fundamental input frequency left peak (yellow
arrow above). The power level
that you record here will be the output power level
(Pout) of the amplifier at the fundamental frequency ( Q2 )
- The gain of the amplifier is Pout-Pin, where Pout is from
question Q2 and Pin is from question Q1. What is the
gain in dB? ( Q3 )
- The two third-order distortion frequencies are clearly seen
at the left and right of the original input frequencies.
- Record the power level, P3 in dBm, and frequency (198.5 MHz)
of the left peak (red arrow above). ( Q4 )
- Compute the output third-order intercept point OIP3 from the
formula in class: OIP3=Pout + (Pout-P3)/2. ( Q5 )
- As discussed in class, if you increase the input power
levels by 10 dB, the third order distortion should increase by
30 dB
- Increase the input level from -25 dBm to -15 dBm, and you
should see increased distortion similar to the following
- Make sure that the marker readout is visible, and make sure
that you can clearly see the power level and frequency of the
peak at the yellow arrow above, then take a photograph for
your report ( P3 )
- Record the power level, P3 in dBm, and frequency (198.5 MHz)
of the left peak (yellow arrow above). ( Q6 )
- How many dB did the level increase from Q4 to Q6? ( Q7 )
- You should not need to use the
following equipment setup information
- Detailed instrument setup instructions, if instructor is not
available
- First ask the instructor to assist, since minor
adjustments are often all that is needed
- Otherwise:
- For the Agilent E4438C Vector Signal Generator to create a
2-tone signal for measurement of OIP3.
- Press “Preset.”
- Press “Frequency” 200 MHz.
- Press Mode>More>Multitone>InitializeTable, and
select NumberTones=2, FreqSpacing=1MHz, Phase=Fixed,
Seed=fixed.
- Press “Mod” on, press “RF” on.
- Press “Amplitude” -20 dBm, “Frequency” 200 MHz, and
Mode>more>multitone>mutitoneOn and ApplyMultitone
- For the Agilent N1996A spectrum analyzer
- Press “Frequency”=200MHz, “span”=10MHz,
“Amplitude”=10dBm.
- You should see two tones at 199.5 and 200.5 MHz from the
signal generator.
- Press “Marker,” and “PeakSearch,” and Marker>more
Add a second marker
- Test Setup/Station 2
- The test setup for test station 2 is shown below
- The top instrument is an Agilent
E4402B spectrum analyzer
- The bottom instruments are a pair of Racal-Dana signal
generators
- First, connect the cable directly from the output of the
signal-combiner (connecting the two signal generators) into
the spectrum analyzer to check the frequencies and power
levels of the two input signal frequencies.
- Note the
signal-combiner in the photo above is used to add the outputs
of the 2 signal generators
- Second, set the 2 signal generatorsfor -20 dBm on the dials.
This requires the dials below the meters to be set to
-20 dBm and the vernier dial adjusted for the meter to read
"0" on the red scale as below:
- You shuld observe two frequencies at 199.5 and 200.5 MHz,
each one at a power level of -24 dBm (24 dB below 1 milliwatt)
on the spectrum analyzer (red arrow below).
- If the signals are already present on the spectrum analyzer
screen, only minor adjustments should be needed.
If the instructor is not available, you may need to refer to
the detailed setup instructions at the end of this document
- Make sure that the signal generator is set for -20 dBm, as
in the picture above
- Then you should observe a spectrum similar to the following
- Press "Marker" and "PeakSearch" to set a marker on the left
spectral line (red arrow above)
- The power level should be approximately -24 dBm at 199.5
MHz,
- Record this power level, Pin in dBm, and frequency (199.5
MHz) of the left peak (red arrow above). The power level that you record here will
be the input power level (Pin) to the amplifier that you will
measure later( Q8 )
- Make sure that the marker readout is visible, and you can
clearly see the power level and frequency of the peak at the
red arrow above, then take a photograph for your report ( P4 )
- Connect the ERA-5 as shown below, with the spectrum analyzer
connected to the output and with the signal generator
connected to the input and 10Vdc power (red=positive,
black=negative)
- See the photo above for use
of the signal-combiner to sum the outputs of the 2 signal
generators
- Recheck to make sure that the
signal generator is set for -20 dBm
- You should observe an output
spectrum similar to the following
- Make sure that the reference
level on this spectrum analyzer is set to 20 dBm
(yellow circle above)
- Make sure that the marker readout is visible, and you can
clearly see the power level and frequency of the peak at the
red arrow above, then take a photograph for your report ( P5 )
- Record the power level, Pout in dBm, and frequency (199.5
MHz) of the fundamental input frequency left peak (yellow
arrow above). The power level
that you record here will be the output power level
(Pout) of the amplifier at the fundamental frequency ( Q9 )
- The gain of the amplifier is Pout-Pin, where Pout is from
question Q9 and Pin is from question Q8. What is the
gain in dB? ( Q10 )
- The two third-order distortion frequencies are clearly seen
at the left and right of the original input frequencies.
- Record the power level, P3 in dBm, and frequency (198.5 MHz)
of the left peak (red arrow above). ( Q11 )
- Compute the output third-order intercept point OIP3 from the
formula in class: OIP3=Pout + (Pout-P3)/2. ( Q12 )
- As discussed in class, if you increase the input power
levels by 10 dB, the third order distortion should increase by
30 dB
- Increase the input level from -20 dBm on the racal-dana
generator's dial to -10 dBm, and you should see increased
distortion similar to the following
- Make sure that the marker readout is visible, and make sure
that you can clearly see the power level and frequency of the
peak at the yellow arrow above, then take a photograph for
your report ( P6 )
- Record the power level, P3 in dBm, and frequency (198.5 MHz)
of the left peak (yellow arrow above). ( Q13 )
- How many dB did the level increase from Q11 to Q13? ( Q14 )
- You should not need to use the
following equipment setup information
- Detailed instrument setup instructions, if instructor is not
available
- First ask the instructor to assist, since minor
adjustments are often all that is needed
- Otherwise:
- For the Racal-Dana generators, see the photographs and
instructions above
- See the photo above for use of the signal-combiner to sum
the outputs of the 2 signal generators
- For the Agilent E4402B spectrum analyzer
- Press “Frequency”=200MHz, “span”=10MHz,
“Amplitude”=10dBm.
- You should see two tones at 199.5 and 200.5 MHz from the
signal generator.
- Press “Marker,” and “PeakSearch,” and Marker>more
Add a second marker
NOTE ReportTemplate: Use the Project Report Template
and keep answers to questions on
consecutive sheets of paper with all plots at the end.
Do not add extraneous pages or put explanations on separate
pages unless specifically directed to do so. The instructor will
not read extraneous pages!
Only turn in requested plots (Pxx )
and requested answers to questions (Qxx ).
All plots must be labeled P1, P2, etc. and all questions must be
numbered Q1, Q2, etc. YOU MUST ADD CAPTIONS AND FIGURE
NUMBERS TO ALL FIGURES!!
Copyright 2010-2015 T. Weldon
Cadence, Spectre and Virtuoso are registered trademarks of
Cadence Design Systems, Inc., 2655 Seely Avenue, San Jose, CA
95134. Agilent and ADS are registered trademarks of Agilent
Technologies, Inc.